YCu-2331
YCu-2331 Automatic Online Monitor of Copper in Water
In the ammonia solution (pH=8锝?0), copper reacts with the chromogenic agent to form a yellow-brown complex: this complex can be extracted with carbon tetrachloride or chloroform, and the absorbance is measured at a specific wavelength. The color is s鈥?/p> Order Online
In the ammonia solution (pH=8锝?0), copper reacts with the chromogenic agent to form a yellow-brown complex: this complex can be extracted with carbon tetrachloride or chloroform, and the absorbance is measured at a specific wavelength. The color is stable for 1h.
In the ammonia solution (pH=8锝?0), copper reacts with the chromogenic agent to form a yellow-brown complex: this complex can be extracted with carbon tetrachloride or chloroform, and the absorbance is measured at a specific wavelength. The color is stable for 1h.
Characteristics:
l Valve components selection. It can select the reagent sampling sequence, with flexible channels, variable functions, minimum dead volume, easy maintenance, long service life and other advantages.
l Printing function. It can be connected to a printer (optional order) to print the measured data in time.
l It adopts 7-inch touch screen, with simple operation interface, easy for users to learn, operate, and maintain.
l Super large data storage capacity. It can realize storage of historical data for no less than 5 years (measurement interval is 1 time/1h), which greatly meets the needs of customers.
l Automatic leakage alarm function. When reagent leakage occurs, the instrument will automatically alarm to remind the user to perform maintenance.
l Light signal recognition function to ensure high accuracy of quantification.
l Simple and convenient chemical reagents maintenance. The reagents can be replaced once a month, which greatly reduces the maintenance of the user.
l U disk one-click upgrade. The software update is much easier.
1 Method basis: spectrophotometry.
2 Measuring range: 0锝?0.0mg/L.
3 Accuracy: 卤10% or 卤0.1mg/L (the larger of the two).
4 Repeatability: 10% or 0.1mg/L.
5 Measurement period: The minimum measurement period is 30 minutes.
6 Sampling cycle: time interval (10锝?999min arbitrarily adjustable) and hourly measurement mode.
7 Calibration cycle: It can be calibrated manually immediately or set cycle for automatic calibration.
8 Maintenance cycle: generally, once a month, each time is about 30 minutes.
9 Output: RS-232.
10 Environment requirements: temperature +5锝?0鈩傷 humidity 鈮?0%RH (no condensation).
11 Power supply: AC (220卤10%) V, (50卤10%) Hz, 5A.